Practical-instrument-electronics 525 Rev B Manual de usuario Pagina 6

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 8
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 5
Model 525 Operating Instructions
82 E. Main Street Suite 3.14 Webster, NY 14580
Tel: 585-872-9350 Fax: 585-872-2638
sales@piecal.com http://www.piecal.com/
Practical Instrument Electronics, Inc. Copyright 2011. All rights reserved.
525-9002 7/13/11
6-8
RTD Temperature Accuracy
The following charts give worst-case temperature accuracy based on stated resistance accuracy of ± (0.015 % + 0.05) Ω.
Pt100 TCR=0.00385
-200 to 850 °C
0.1
0.2
0.3
0.4
-200 0 200 400 600 800
Temperature °C
Error °C
Pt100 TCR=0.003902
-200 to 648.9 °C
0.1
0.2
0.3
0.4
-200 0 200 400 600 800
Temperature °C
Error °C
Pt100 TCR=0.003916
-200 to 648.9 °C
0.1
0.2
0.3
0.4
-200 0 200 400 600 800
Temperature °C
Error °C
Pt100 TCR=0.003926
-200 to 850 °C
0.1
0.2
0.3
0.4
-200 0 200 400 600 800
Temperature °C
Error °C
Cu10 TCR=0.00427
-200 to 260 °C
1.2
1.25
1.3
1.35
1.4
-200 -100 0 100 200 300
Temperature °C
Error °C
Cu50 TCR=0.00428
-50 to 150 °C
0
0.1
0.2
0.3
0.4
0.5
-50 0 50 100 150
Temperature °C
Error °C
Ni120 TCR=0.00672
-80 to 260 °C
0.07
0.08
0.09
0.1
-100 -50 0 50 100 150 200 250 300
Temperature °C
Error °C
Ni110 TCR=0.00530
-100 to 260 °C
0.08
0.1
0.12
0.14
0.16
-100
0
100
200
300
400
Temperature °C
Error °C
Vista de pagina 5
1 2 3 4 5 6 7 8

Comentarios a estos manuales