Product Description
• EASY TO USE
With the PIECAL Model 334 you can check, calibrate
and measure all your current signal instruments in a
4 to 20 milliamp DC loop. It can be used at any
access point in your loop. Source & Read 0.00 to
24.00 mA, Simulate a 2 Wire Transmitter or use the
PIECAL Model 334 to simultaneously power your 2
Wire Transmitter and measure its output. When
desired the PIECAL Model 334 can display current
in milliamps or percent of 4 to 20.
• SOURCE MILLIAMPS
Calibrate recorders, digital indicators, stroke valves
or any instruments that get their input from a 4 to
20 mA loop. Easily set any value quickly to within
0.01 mA with the adjustable digital potentiometer
“DIAL” or use preset 4.00 mA (0.0%) and 20.00
mA (100.0%) EZ-CHECK™ settings.
• RECALL OUTPUT SETTINGS
The EZ CHECK™ switch provides rapid checking of
4.00, 20.00 and any convenient third point between
0.00 to 24.00 mA.
• CALIBRATE USING LOOP POWER
Check loop wiring and receivers by using the
PIECAL Model 334 in place of a 2 Wire transmitter.
Simulate a changing process input to check loop
response and control settings. The PIECAL Model
334 uses any loop power from 2 to 100V DC.
• READ LOOP CURRENT
Check controller outputs or measure the milliamp signal anywhere in the loop. The
PIECAL Model 334 measures 0.00 to 52.00 mA signals with greater accuracy than
a typical multimeter. The PIECAL Model 334 can be easily switched to display
milliamps or percent of 4 to 20.
• POWER & MEASURE 2 WIRE TRANSMITTERS
The PIECAL Model 334 can simultaneously output 24V DC to power any and all devices in a
process loop using the internal batteries and the internal switching power supply, while
measuring the output of a 2 Wire Transmitter and any other loop devices. This is handy for
checking the functionality of transmitters in the field or on the bench.
• READ DC VOLTS
The PIECAL Model 334 can measure from -99.99 to +99.99 VDC with 10mV resolution. Use it to
check loop power supplies, I/V converters, chart recorders, 1 to 5 Volt signals, and any other
voltages within this range making it unnecessary to carry an additional multimeter.
PIECAL Model 334
4-20 mA Loop Calibrator
Operating Instructions
PIECAL Model 334
4-20 mA Loop Calibrator
Operating Instructions
Practical Instrument Electronics
82 East Main Street Suite 3.14 • Webster, NY 14580 Tel: 585.872.9350 • Fax: 585.872.2638 • [email protected] • www.piecal.com
Copyright © 2011 All rights reserved • 334-9002 Rev E - 13 July 2011
Vista de pagina 0
1 2 3 4 5 6 7 8

Indice de contenidos

Pagina 1 - PIECAL Model 334

Product Description• EASY TO USEWith the PIECAL Model 334 you can check, calibrate and measure all your current signal instruments in a 4 to 20 milli

Pagina 2 - Basic Operation

q POWER SWITCHSelect “mA” to display and calibrate in milliamps.Select “% 4 to 20 mA” to display and calibrate in percent. Select “READ VDC” to read

Pagina 3 - Reading Milliamp Outputs

mA OUT, % OUT (Percent of 4 to 20 mA)Choose this function to provide an output from 0.00 to 24.00 milliamps. The compliance voltage is a nominal 24 VD

Pagina 4 - Simulate 2-Wire Transmitters

2 Wire mA, 2 Wire % (Percent of 4 to 20 mA)Choose this function to simulate a 2 Wire Transmitter output from 0.00 to 24.00 milliamps. Operates in loop

Pagina 5 - Application Notes

READ VChoose this function to measure from -99.99 to +99.99V DC.1) Select “READ VDC” with slide switch q.2) Connect the red (+) and black (-) leads of

Pagina 6 - Guarantee

AccessoriesAC ADAPTOR (200 to 240 VAC) Part No. 020-0100AC ADAPTOR (100 to 120 VAC) Part No. 020-0101Ni-MH 1 Hour Charger w/4 Ni-MH AA Batte

Pagina 7 - PIECAL 334 Specifications

GeneralOperating Temperature Range -20 to 60 °C (-5 to 140 °F)Storage Temperature Range -30 to 60 °C (-22 to 140 °F)Relative Humidity Range 10 % ≤RH ≤

Pagina 8 - Additional Information

Source/Power & Measure Two Wire Transmitters334 Ranges and Resolution0.00 to 24.00 mA Full Span or -25.0 to 125.0% of 4-20 mAAccuracyEZ Check(s)